Probe and Sensor Node Configuration
1 bi-directional, dual 1/2� NPT female bleed sensor housing
Installed Accuracy
Airflow through an opening or across and obstruction: Requires
field measurement of a reference airflow of the specific installation.
The Field Adjust Wizard (FAW) facilitates setup.
Equivalent pressure between two adjacent spaces: Requires field
measurement of a reference pressure to correct the default flow
coefficient of the specific installation. The Field Adjust Wizard (FAW)
facilitates setup.
Listings and Compliance
UL: 60730-1, 60730-2-9; CAN E60730-1, E60730-2-9 (EF-A2000-B
Only)
FCC: This device complies with Part 15 of the FCC rules
RoHS: This device is RoHS2 compliant
Environmental Limits
Temperature:
Sensor -2,000 to 2,000 fpm [-10.16 to 10.16 m/s]:
-20 to 160 ?F [-28.9 to 71.1 ?C]
Sensor -3,000 to 3,000 fpm [-15.24 to 15.24 m/s]:
0 to 160 ?F [-17.8 to 71.1 ?C]
Transmitter: -20 to 120 ?F [-28.9 to 48.9 C]
Humidity: (non-condensing)
Probes: 0 to 100%
Transmitter: 5 to 95%